BOSTON SCIENTIFIC CORPORATION ADVANIX PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00537300
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix pancreatic stent was to be used during a pancreatic stent placement procedure in the pancreatic duct, performed on (b)(6) 2022.During preparation, it was noted that the stent was kinked and could not be mounted into the delivery system.Another advanix pancreatic stent was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: medical device problem code a0406 captures the reportable event of pancreatic stent kinked.Block h10: the returned advanix pancreatic stent was analyzed, and a visual evaluation noted that there were kinks in the middle section of the stent which resulted to problems or difficulty to prepare the device.No other problems with the device were noted.The reported event was confirmed.Based on all the collected information and product analysis, the device met all manufacturing requirements, and no abnormalities were reported during the manufacturing process.The problem of stent kinked occurred during preparation, it is possible that operational factors, such as user technique/handling, contributed to the observed stent kinks problem, this condition could have affected the device integrity.Therefore, adverse event related to procedure is selected as the most probable root cause for the complaint.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix pancreatic stent was to be used during a pancreatic stent placement procedure in the pancreatic duct, performed on (b)(6), 2022.During preparation, it was noted that the stent was kinked and could not be mounted into the delivery system.Another advanix pancreatic stent was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX PANCREATIC STENT
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15761918
Report Number3005099803-2022-06443
Device Sequence Number8221091
Product Code FGE
UDI-Device Identifier08714729856894
UDI-Public08714729856894
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133700
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2023
Device Model NumberM00537300
Device Catalogue Number3730
Device Lot Number0028244679
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-