BIOMET 3I 3I T3¿ WITH DCD¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 5 X 8.5MM; DENTAL IMPLANT Back to Search Results
Model Number BNST585
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Pain (1994); Osteolysis (2377)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported implant removed due to pain.Bone graft was placed time of implant removal.New implant was placed on (b)(6) 2022.Uncovering was also done on (b)(6) 2022.Tooth 19.Symptoms as a result of the event: pain, inflammation.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient weight unknown / not provided.Date of event unknown / not provided.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3¿ WITH DCD¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 5 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15762106
Report Number0001038806-2022-01719
Device Sequence Number8221584
Product Code DZE
UDI-Device Identifier00844868029335
UDI-Public(01)00844868029335(17)250227(10)2020020864
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/27/2025
Device Model NumberBNST585
Device Catalogue NumberBNST585
Device Lot Number2020020864
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2022
Initial Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 YR
Patient SexFemale
-
-