| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Will be returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the grey tip impactor broke during impaction.Attempts have been made and there is no further information at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Reported event was confirmed as visual examination of the returned product identified the incorrect tip was returned with the impactor handle.A grey tip was returned with the impactor, when the impactor print calls out a black tip.Both the tip and shaft show signs of wear from use.It is confirmed that the tip has fractured and not all parts were returned.The features of the fracture surface visually align with a bending overload fracture failure.Device history record was reviewed and no discrepancies relevant to the reported event were found.The tips of the impactors switched are epoxied, and therefore not to be disassembled for sterilization.However, it is unknown if this contributed to the reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|