Model Number 0502880430
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was lack of image.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was lack of image.
 
Additional Manufacturer Narrative
Alleged failure: can't see through confirmed failure: outer tube damaged (bent, dented) probable root cause: ¿ incorrectly assembled optical train ¿ damage to optical train ¿ end of life wear-out ¿ shipping damage ¿ use error the product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, SEMI-RIGID URETEROSCOPE, 43CM
Common Device Name
URETEROSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15762638
Report Number0002936485-2022-00640
Device Sequence Number18260976
Product Code FGB
UDI-Device Identifier07613327063592
UDI-Public07613327063592
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K121112
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502880430
Device Catalogue Number0502880430
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/15/2022
Supplement Date Received by Manufacturer10/15/2022
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date01/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-