| Model Number |
E801 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The alarm trace did not show a conspicuous event.The investigation found that the sample was run in a non-standard sample cup.Based on the data provided, a general reagent issue can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys free psa immunoassay results for 1 patient sample on a cobas e 801 analytical unit.The initial free psa result was 6.58 ng/ml.The repeat result was 3.19 ng/ml.The initial result was not reported outside of the laboratory.The repeat result was deemed correct.The reagent lot is 61763201.The expiration date is 30-jun-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration was last performed on 11-jul-2022 and showed signals below the expected range.Qc performed on the day of the event was acceptable.
|
| |
|