| Model Number |
1012447-12 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the mid left anterior descending (lad) with heavy calcification, moderate tortuosity and 99% stenosis.The 2.5x12 mm nc trek balloon dilatation catheter was advanced to the lesion with resistance with the anatomy noted and was inflated once for pre-dilatation and ruptured at 12 atmospheres for 10 seconds.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|