MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERLOC MAX POWER-INJECTIBLE INFUSION SET 20G X 1.0 IN; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERLOC MAX POWER-INJECTIBLE INFUSION SET 20G X 1.0 IN; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported unit has been had issue with port needle line cracking.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a crack in the infusion set was inconclusive due to the sample condition.Two photographs of the implicated 20g x 1¿ powerloc max infusion set were returned for evaluation.The y-site of the device was within frame of the photograph while the rest of the device was outside the frame of the photograph.An injection cap was attached to the y-site and an orange cap was partially attached to the injection cap.No cracks or damage were observable in the returned photographs.Microscopic examination and functional testing could not be conducted without the receipt of the physical sample.Therefore, the reported failure mode could not be confirmed, and a specific root cause could not be identified.Possible contributing factors for a leak at the connection could include a cracked luer adaptor or incompatibility with an interfacing device.If the physical sample is returned at a later date, this evaluation will be updated with the relevant information.A lot history review (lhr) of asgsfc059 showed one other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported unit has been had issue with port needle line cracking.No other information was provided.Additional information received 12/22/2022: it was reported by bd sales rep the cracks are occurring at the luer connector, where the needleless injection cap is attached.
 
Event or Problem Description
It was reported unit has been had issue with port needle line cracking.No other information was provided.Additional information received 12/22/2022: it was reported by bd sales rep the cracks are occurring at the luer connector, where the needleless injection cap is attached.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC MAX POWER-INJECTIBLE INFUSION SET 20G X 1.0 IN
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
parque industrial colonial
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15763884
Report Number3006260740-2022-04522
Device Sequence Number20397867
Product Code FPA
UDI-Device Identifier00801741047350
UDI-Public(01)00801741047350
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0132010
Device Lot NumberASGSFC059
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer12/13/2022
01/10/2023
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date12/30/2022
01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-