AVANOS MEDICAL INC. HOMEPUMP C-SERIES, 100 ML, 2 ML/HR (CONTAINS DEHP); ELASTOMERIC LFR Back to Search Results
Model Number C100020
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical received a single report that referenced three different incidences, which were associated with separate units, involving three different events.This is the first of three reports.Refer to 2026095-2022-00120 for the second event.Refer to 2026095-2022-00121 for the third event.Fill volume: 100 ml.Flow rate: unknown.Procedure: chemotherapy.Cathplace: unknown.It was reported there was a reported fast flow event.The clinicians were using nominal fill volumes for medication (chemotherapy) to infuse over 46-hours.The patient returned ten-twelve hours earlier than expected.The user noted once they increased the diluent, the issue "corrected itself and they have not had any further incidents of fast flow.There were no reported symptoms or injuries.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 09-nov-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMEPUMP C-SERIES, 100 ML, 2 ML/HR (CONTAINS DEHP)
Common Device Name
ELASTOMERIC LFR
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c. 22116
MX   22116
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15764875
Report Number2026095-2022-00119
Device Sequence Number15571605
Product Code MEB
UDI-Device Identifier00193494135614
UDI-Public00193494135614
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K052117
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC100020
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Initial Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CHEMOTHERPY
Patient Age65 YR
Patient SexMale
Patient Weight114 KG
-
-