| Model Number |
CFF71 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit is anticipated to return to applied medical for evaluation.A follow-up report will be sent upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: colon op.Event description: when inserting the trocar, the patient's resistance was so strong that the obturator tip was bent.No foreign body or bone was touched.Additional information received from applied medical representative via email 3nov22: the patient is fine.Converted the surgery from laparoscopic to open, which would have been necessary even without the prolapse of the trocars.At that time, no other instruments were used.The trocars were rotated and inserted at an angel (45°).The trocars were not used with a robot.No cannulas were inserted using the obturator prior to the incident.No pieces fell into the patient.The fourth trocar is a cff33, it was not damaged and is not affected.Additional information received from applied medical representative via email 4nov22: the operation should have been converted even without the damage to the trocars.The reason they had to convert the surgery from laparoscopic to open was that the adhesions were too much.The patient had a previous surgery in this area.There was cartilage or scar tissue underneath the insertion area the fourth trocar is not returning.Intervention: converted the surgery from laparoscopic to open, which would have been necessary even without the prolapse of the trocars.Patient status: the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of a damaged obturator tip.The obturator tip was fractured.Based on the description of the event and the condition of the returned unit, it is likely that the obturator tip fractured due to excess force exerted on the tip during insertion due to the adhesions.It is also possible that the obturator material, which is susceptible to fracture, was a contributing factor.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|
Event or Problem Description
|
|
Procedure performed: colon op.Event description: when inserting the trocar, the patient's resistance was so strong that the obturator tip was bent.No foreign body or bone was touched.Additional information received from applied medical representative via email 3nov22: the patient is fine.Converted the surgery from laparoscopic to open, which would have been necessary even without the prolapse of the trocars.At that time, no other instruments were used.The trocars were rotated and inserted at an angel (45°).The trocars were not used with a robot.No cannulas were inserted using the obturator prior to the incident.No pieces fell into the patient.The fourth trocar is a cff33, it was not damaged and is not affected.Additional information received from applied medical representative via email 4nov22: the operation should have been converted even without the damage to the trocars.The reason they had to convert the surgery from laparoscopic to open was that the adhesions were too much.The patient had a previous surgery in this area.There was cartilage or scar tissue underneath the insertion area.The fourth trocar is not returning.Intervention: converted the surgery from laparoscopic to open, which would have been necessary even without the prolapse of the trocars.Patient status: the patient is fine.
|
| |
|