• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS TURBO-POWER LASER ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS TURBO-POWER LASER ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 423-050
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thromboembolism (2654)
Date of Event 08/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated by mfr: the device was discarded, thus no investigation could be completed.Although embolism is a known risk of complication with use of the turbo-power device, the physician stated that there was pre-existing thrombus within the stents prior to the turbo-power being used and expected a potential embolization.Investigation determined this event to be use error, for not using mechanical thrombectomy prior to utilization of the turbo-power.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
As part of a japan turbo postmarket study, a peripheral atherectomy procedure began to treat an in-stent restenosis (isr) in the patient's mid superficial femoral artery (sfa) to the popliteal region.The lesion to be treated was 400 mm in length, with heavy thrombus present within the stents.It was predicted that treatment would result in clots and potential embolism; therefore, a filter was placed pre-procedure, distal to the lesion.The physician chose to treat the patient with a spectranetics turbo-power laser atherectomy catheter.During use, and as expected, portions of the thrombus were observed.The filter did not catch all the thrombus, resulting in a peripheral embolism.The thrombus that was not captured by the filter was removed successfully using a mechanical thrombectomy catheter.The procedure was completed with no reported patient harm.This report captures the turbo-power in use within the lesion when the peripheral embolism occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TURBO-POWER LASER ATHERECTOMY CATHETER
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key15765107
Report Number1721279-2022-00193
Device Sequence Number6353508
Product Code MCW
UDI-Device Identifier00813132026776
UDI-Public(01)00813132026776(17)240523(10)FMP22E06A
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K180694
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/23/2024
Device Model Number423-050
Device Catalogue Number423-050
Device Lot NumberFMP22E06A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2022
Initial Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER MANUFACTURER AND SIZE UNK.; GUIDE WIRE MANUFACTURER AND SIZE UNK.; INTRODUCER SHEATH MANUFACTURER AND SIZE UNK.; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight49 KG
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-