Model Number G6626526
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a distributor via manufacturer representative regarding cages used in spinal therapy.It was reported that when the surgery began, the doctor removed the entire disc and prepared the space.He used the gauge to verify the height of the divergence box, and a box with a height of 6mm was placed on the table since it was the measure chosen with the divergence gauge height.It was placed in the freehand inserter and filled with graft, the doctor then placed it in the space and released it from the inserter.He then impacted a little to finish locating it, using the impactor and the hammer, the cage fractured where the locking screw is located.The surgeon removed the box and then decided to replace it by a 5mm high divergence cage.When the 5mm high divergence box was passed to the table, in the same way it was placed in the freehand inserter and filled with graft.Once again, the doctor placed the box, it entered the space much more smoothly than the previous one since it was a lower height, but in the same way when impacting it a little to finish accommodating it, the cage fractured.No patient symptoms were reported.No further complications were reported/anticipated.
 
Additional Manufacturer Narrative
H1: review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunction.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Mdr decision corrected to not reportable.No additional supplementals required unless additional information received indicates reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIVERGENCE¿ ANTERIOR CERVICAL FUSION SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED F
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key15768507
Report Number1030489-2022-01024
Device Sequence Number14695587
Product Code OVE
UDI-Device Identifier00643169337497
UDI-Public00643169337497
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K141599
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG6626526
Device Catalogue NumberG6626526
Device Lot NumberH5785573
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer11/25/2022
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date11/28/2022
Date Device Manufactured08/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
-
-