As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.Expiry date: 09/2024.
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one seeker crossing support catheter was returned for evaluation.During visual evaluation, the tip was noted to be detached.Segment measures 149 cm from break to strain relief.No functional testing due to the condition of the device.One electronic photo was provided and reviewed.The photo shows the seeker package and label of the device.The product is not clearly visible in the provided photo.Therefore, based on the photo review, the investigation can not be confirmed for the reported detachment issue.One fluoroscopic image was provided and reviewed.The image showed the right thigh.Part of a wire is seen likely in the proximal superficial femoral artery.Distal to this is an object that resembles a piece of the catheter.The remaining portion of the catheter is not visualized.Therefore, based on the image review, the investigation is confirmed for the reported detachment issue.Therefore, the investigation is confirmed for detachment of device or device component issue as the tip was noted to be detached.A definitive root cause for the reported detachment issue could not be determined based upon the provided information.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: b5, d4 (expiry date: 09/2024), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|