Model Number SK15035M
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.Expiry date: 09/2024.
 
Event or Problem Description
It was reported that during a recanalization procedure, the tip of the device was allegedly detached.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one seeker crossing support catheter was returned for evaluation.During visual evaluation, the tip was noted to be detached.Segment measures 149 cm from break to strain relief.No functional testing due to the condition of the device.One electronic photo was provided and reviewed.The photo shows the seeker package and label of the device.The product is not clearly visible in the provided photo.Therefore, based on the photo review, the investigation can not be confirmed for the reported detachment issue.One fluoroscopic image was provided and reviewed.The image showed the right thigh.Part of a wire is seen likely in the proximal superficial femoral artery.Distal to this is an object that resembles a piece of the catheter.The remaining portion of the catheter is not visualized.Therefore, based on the image review, the investigation is confirmed for the reported detachment issue.Therefore, the investigation is confirmed for detachment of device or device component issue as the tip was noted to be detached.A definitive root cause for the reported detachment issue could not be determined based upon the provided information.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: b5, d4 (expiry date: 09/2024), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure, the tip of the device was allegedly detached.It was further reported that the detached part could not be retrieved by using snare.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEEKER
Common Device Name
SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15769091
Report Number2020394-2022-00887
Device Sequence Number15075531
Product Code DQY
UDI-Device Identifier00801741067921
UDI-Public(01)00801741067921
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K103367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSK15035M
Device Catalogue NumberSK15035M
Device Lot NumberVTGW0047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date01/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-