Model Number 7K62-35
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported a falsely decreased architect tsh result on a 9-yr old female patient.Results provided: (b)(6) 2022, sid (b)(6) = 85.94 / > 100 / x5 dilution = 190.03 / 197.66 uiu/ml no impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.Patient identifier complete sid: (b)(6).
 
Additional Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.Section a1 patient identifier sids: (b)(6).Additional testing added to section b5: new sample procured on the same patient on 15nov2022: sid: (b)(6).Tsh 53.83 uiu/ml.Ft3 < 1.50 pg/ml.Ft4 0.54 ng/dl.Repeated on 11/16/2022.Ft4 < 0.40 ng/dl.Tsh > 100.00 uiu/ml.Ft3 < 1.50 pg/ml.Tsh 1:5 dilution = 116.02 uiu/ml.
 
Event or Problem Description
The customer reported a falsely decreased architect tsh result on a 9-yr old female patient.Results provided: 26oct2022 sid (b)(6) = 85.94 / > 100 / x5 dilution = 190.03 / 197.66 uiu/ml.New sample procured on the same patient on 15nov2022: sid: (b)(6).Tsh 53.83 uiu/ml.Ft3 < 1.50 pg/ml.Ft4 0.54 ng/dl.Repeated on 11/16/2022.Ft4 < 0.40 ng/dl.Tsh > 100.00 uiu/ml.Ft3 < 1.50 pg/ml.Tsh 1:5 dilution = 116.02 uiu/ml no impact to patient management was reported.
 
Additional Manufacturer Narrative
The evaluation of complaint data for the product and likely cause architect tsh reagent lot 39336ud00 identified normal complaint activity and there are no trends for the product related to patient results.No customer returns were available for evaluation.Historical performance in the field of reagent lot using worldwide data through abbottlink was evaluated.The median population result for the lot is within established baselines and comparable with all other lots in the field and confirms no systemic issue for this lot.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lot.This review did not identify any non-conformances, potential non-conformances or deviations.A review of labeling concluded that the issue is sufficiently addressed.Based on our investigation no systemic issue or product deficiency of the architect tsh reagent lot 39336ud00 was identified.
 
Event or Problem Description
The customer reported a falsely decreased architect tsh result on a 9-yr old female patient.Results provided: 26oct2022 sid (b)(6) = 85.94 / > 100 / x5 dilution = 190.03 / 197.66 uiu/ml.New sample procured on the same patient on 15nov2022: sid: (b)(6).Tsh 53.83 uiu/ml.Ft3 < 1.50 pg/ml.Ft4 0.54 ng/dl.Repeated on 11/16/2022.Ft4 < 0.40 ng/dl.Tsh > 100.00 uiu/ml.Ft3 < 1.50 pg/ml.Tsh 1:5 dilution = 116.02 uiu/ml no impact to patient management was reported.
 
Additional Manufacturer Narrative
The evaluation of complaint data for the product and likely cause architect tsh reagent lot 39336ud00 identified normal complaint activity and there are no trends for the product related to patient results.No customer returns were available for evaluation.Historical performance in the field of reagent lot using worldwide data through abbottlink was evaluated.The median population result for the lot is within established baselines and comparable with all other lots in the field and confirms no systemic issue for this lot.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lot.This review did not identify any non-conformances, potential non-conformances or deviations.A review of labeling concluded that the issue is sufficiently addressed.Based on our investigation no systemic issue or product deficiency of the architect tsh reagent lot 39336ud00 was identified.Section a1 patient identifier sids: (b)(6).Additional testing added to section b5: new sample procured on the same patient on 15nov2022: sid: (b)(6).Tsh 53.83 uiu/ml.Ft3 < 1.50 pg/ml.Ft4 0.54 ng/dl.Repeated on 11/16/2022.Ft4 < 0.40 ng/dl.Tsh > 100.00 uiu/ml.Ft3 < 1.50 pg/ml.Tsh 1:5 dilution = 116.02 uiu/ml.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TSH REAGENT KIT
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15769333
Report Number3005094123-2022-00249
Device Sequence Number17429469
Product Code JLW
UDI-Device Identifier00380740105969
UDI-Public00380740105969
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K983442
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/29/2023
Device Model Number7K62-35
Device Catalogue Number07K62-35
Device Lot Number39336UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/28/2022
Supplement Date Received by Manufacturer11/17/2022
01/09/2023
02/08/2023
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date12/08/2022
01/10/2023
02/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/09/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2K PROC MOD, 03M74-01, ISR60829; ARC I2K PROC MOD, 03M74-01, ISR60829; ARC I2K PROC MOD, 03M74-01, ISR60829; ARC I2K PROC MOD, 03M74-01, ISR60829
Patient Age9 YR
Patient SexFemale
-
-