RESMED LTD AIRSENSE 11 AUTOSET USA TRI; Ventilator, non-continuous (respirator) Back to Search Results
Model Number 39001
Medical Device Problem Code Unintended Electrical Shock (4018)
Health Effect - Clinical Codes Burn(s) (1757); Chest Pain (1776); Dysphagia/ Odynophagia (1815); Pain (1994); Electric Shock (2554); Unspecified Heart Problem (4454)
Date of Event 10/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to resmed that a patient was allegedly electrocuted by his airsense 11 device during use.The patient bore burn injuries.The patient sought medical attention in an emergency care unit.The patient is currently still experiencing pain in chest, heart issues and trouble swallowing following the incident.Further information about the patient¿s medical condition is not available to resmed.
 
Additional Manufacturer Narrative
The airsense 11 device was returned to resmed.The investigation methods, results, and conclusions are not finalised at this stage.If further information becomes available, a supplementary report will be submitted.The airsense 11 user guide states the following warning:- ¿beware of electrocution: do not immerse the device, ac adaptor or power cord in water.Do not connect to power while the device is wet.Make sure that all parts are dry before plugging it in.If liquids are spilled into or onto the device, unplug the device and let the parts dry.¿ the user guide also provide the following warning:- ¿discontinue use and contact your care provider or resmed service center if any of the following occur: device does not perform as usual device is making unusual sounds device is damaged.¿ resmed reference#: (b)(4).Pending investigation.
 
Additional Manufacturer Narrative
The airsense 11 was returned to resmed for investigation.Performance testing revealed the device passed all functional tests and the device was performing per specification.The investigation concluded there was no relationship between use of the device and the reported incident.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event or Problem Description
It was reported to resmed that a patient was allegedly electrocuted by his airsense 11 device during use.The patient bore burn injuries.The patient sought medical attention in an emergency care unit.The patient is currently still experiencing pain in chest, heart issues and trouble swallowing following the incident.Further information about the patient¿s medical condition is not available to resmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRSENSE 11 AUTOSET USA TRI
Common Device Name
Ventilator, non-continuous (respirator)
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key15769398
Report Number3004604967-2022-00753
Device Sequence Number12417228
Product Code BZD
UDI-Device Identifier00619498390014
UDI-Public(01)00619498390014(10)1616759
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K203126
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number39001
Device Catalogue Number39001
Device Lot Number1616759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date03/21/2023
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer03/21/2023
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/18/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient Age34 YR
Patient SexMale
Patient Weight180 KG
-
-