SYNTHES GMBH 5.0 TI LCKNG SCR SLF-TPNG T25 SD 44-SILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 412.216S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Pain (1994); Post Operative Wound Infection (2446); Thrombosis/Thrombus (4440); Swelling/ Edema (4577)
Type of Reportable Event Death
Event or Problem Description
It was reported that on (b)(6) 2021 the patient underwent right hip hemiarthroplasty and removal of the hardware.Upon inspection of the hardware the depuy synthes femoral neck system had broken completely through the plate.The patient struggled from this procedure and suffered significant swelling in his bilateral lower extremities which was caused by deep vein thrombosis in his popliteal vein.The patient subsequently developed an infection which required a third procedure to remove the hardware and put in an antibiotic spacer on (b)(6) 2021.On (b)(6) 2021 the patient died due to cardiac arrest, anemia post op, septic right hip and thrombosis of the left arm and right leg.On or about december 08, 2020 the patient suffered a fall at his home, while carrying a few gallons of water on his property which included a sloping driveway.X-ray reported a right intertrochanteric fracture and a ct which concluded that he had impacted comminuted right intertrochanteric fracture.On (b)(6) 2020 the patient underwent open treatment of the right femoral neck fracture with a screw and side plate.On (b)(6) 2021 the patient complaints of worsening right hip pain and lower extremity edema.Stated the pain was so bad, he was unable to stand, the pain being worse along the inguinal crease.On (b)(6) 2021 he reiterated his worsening pain and had to ambulate with a walker.From (b)(6) 2021 the patient was back at the hospital with right hip pain that was constant.He stated that he had suffered no falls or injuries since the surgery.During his 3-day stay, was given x-rays, an mri of his lumbar spine without contrast and had a fluoroscopy and epidurogram with steroid injection at l2-l3.He was then discharged to a skilled nursing facility.On (b)(6) 2021 the patient returned to the healthcare facility where he complained of continued pain and trouble ambulating.He also stated that his surgical wound had an odor.He returned to the rehab center where he remained for a month until he suffered a fall.On (b)(6) 2021 he was found to have an infected right total hip arthroplasty, surgical wound dehiscence, a uti, acute metabolic encephalopathy and a stage iii pressure ulcer on his right heel.This report involves one (1) 5.0mm ti locking scr slf-tpng w/t25 stardrive 44mm-sterile.This is report 1 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the 5.0 ti lckng scr slf-tpng t25 sd 44-sile appears assembled to a broken fragment of femoral neck system pl 2 hole ¿ sterile.Based on the evidence provided, no signs of breakage were observed on the surface of the device.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for 5.0 ti lckng scr slf-tpng t25 sd 44-sile.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot part number: 412.216s-us, lot number: 32p4097, manufacturing site: mezzovico , release to warehouse date: 14 jan 2020, expiration date: 31 dec 2029.A manufacturing record evaluation was performed for the sterile finished lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 TI LCKNG SCR SLF-TPNG T25 SD 44-SILE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK MEZZOVICO (CH)
via cavazz 5
mezzovico 6805
SZ   6805
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15769648
Report Number8030965-2022-09567
Device Sequence Number6620612
Product Code KTT
UDI-Device Identifier10886982275431
UDI-Public(01)10886982275431(17)291231(10)32P4097
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K000682
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number412.216S
Device Catalogue Number412.216S
Device Lot Number32P4097
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/18/2022
Supplement Date Received by Manufacturer11/16/2022
12/15/2022
04/05/2023
07/07/2023
08/23/2024
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date12/08/2022
12/16/2022
04/21/2023
07/25/2023
09/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.0 TI LCKNG SCR SLF-TPNG T25 SD 44-SILE.; ANTIROTATION SCR FEM NECK 105 LNTH - S.; BOLT FOR FEMORAL NECK SYS 105 LENGTH-S.; FEMORAL NECK SYSTEM PL 2 HOLE - STERILE.
Outcome Attributed to Adverse Event Required Intervention; Death;
Patient Age83 YR
Patient SexMale
-
-