| Model Number |
TRULIANT TIB IMP PS INSERT SZ 2.5 9MM |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a surgical procedure, on the back table, a black l shape object was found on the truliant insert.This object was in the evoh package.The surgeons would like to know where this object came from.No impact to a patient since this was found before surgery.Surgical delay/prolongation, less than 5 minutes.Pcs with same size insert was implanted.The device is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
The black foreign matter within sterile packaging reported may have been due to either variance in improper maintenance of ultrasonic/alcohol black polypropylene drying mats, lack of operator awareness during visual inspection, inconsistency in red zone disinfection frequency, and/or inadequately defined interim product storage/cleaning maintenance for blue and red zone packaging areas compromising product sterility of pouch sealed product.Additionally, a lack of routine training decreased awareness around contamination complaints historically investigated.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|