Model Number I400+
Medical Device Problem Codes Incorrect Measurement (1383); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for three patient samples tested with the na+ electrode on a cobas integra 400 plus.No incorrect results were reported outside of the laboratory.The first sample initially resulted in a na value of 126 mmol/l and repeated with a value of 132 mmol/l.The second sample initially resulted in a na value of 118 mmol/l and repeated with a value of 138 mmol/l.The third sample initially resulted in a na value of 120 mmol/l and repeated with a value of 140 mmol/l.The na electrode lot number and expiration date were requested, but not provided.
 
Additional Manufacturer Narrative
The field service engineer determined the ise module had an initialization problem.The ise mixing tower and a system reagent were changed.The electrode block and seals were checked.The ise tubing was refitted and verified to be seated correctly into the valves.Calibrator solution was primed and a sensor was checked.Calibration and controls were ok.The issue was corrected after the service actions.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INTEGRA 400 PLUS ANALYZER
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE INSTRUMENT CENTER AG TEGIMENTA
forrenstrasse
na
rotkreuz 6343
SZ   6343
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15770121
Report Number1823260-2022-03570
Device Sequence Number17764813
Product Code JJE
UDI-Device Identifier04015630032457
UDI-Public04015630032457
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberI400+
Device Catalogue Number03245233001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-