| Model Number |
C503 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable alp2 alkaline phosphatase acc.To ifcc gen.2 results for 1 patient sample on a cobas pro c 503 analytical unit.The initial alp2 result was 337 u/l.The repeat result was 182 u/l.The sample was repeated again on another analyzer and the result was 190 u/l.The initial result was not reported outside of the laboratory.The reagent lot number is 654864 and the expiration date is 31-may-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The alarm trace did not contain a conspicuous event.Based on the calibration and qc data, we could rule out a general reagent issue.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The issue was resolved by changing the reagent and recalibrating the assay.The investigation did not identify a product problem.
|
| |
|