| Model Number |
87047 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The catheter was returned to boston scientific for analysis.Upon device analysis visual inspection revealed the irrigation extension tubing was found partially detached/separated from the luer fitting at the adhesive joint.The functional test revealed no abnormal resistance was felt when actuating the steering mechanism.The product analysis confirms the reported complaint.
|
| |
|
Event or Problem Description
|
|
During catheter preparation of a procedure a intellanav mifi open-irrigated was selected for use.The fluid port disconnected from the catheter tubing on product.The procedure was completed successfully with a catheter of the same make and model and the patient had no issues.
|
| |
|
Event or Problem Description
|
|
During catheter preparation of a procedure a intellanav mifi open-irrigated was selected for use.The fluid port disconnected from the catheter tubing on product.The procedure was completed successfully with a catheter of the same make and model and the patient had no issues.
|
| |
|