MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/AJLA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/AJLA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN917286
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that "the patient was prepared in the operating room on (b)(6) 2022 for coronary artery bypass grafting, left ventricular thrombectomy, ventricular wall aneurysm resection and suturing under general anesthesia and hypothermic extracorporeal circulation.Iabp was used to give cardiac assist, the product showed severe ecg interference and could not monitor ecg.The patient was not harmed by the replacement of the iabp".The patient status is reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Additionally no recorder strip was returned for investigation.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the patient was prepared in the operating room on (b)(6) 2022 for coronary artery bypass grafting, left ventricular thrombectomy, ventricular wall aneurysm resection and suturing under general anesthesia and hypothermic extracorporeal circulation.Iabp was used to give cardiac assist, the product showed severe ecg interference and could not monitor ecg.The patient was not harmed by the replacement of the iabp".The patient status is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/AJLA
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15771736
Report Number3010532612-2022-00485
Device Sequence Number6621972
Product Code DSP
UDI-Device Identifier10801902172072
UDI-Public10801902172072
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN917286
Device Catalogue NumberIAP-0701
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer12/13/2022
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age41 YR
Patient SexMale
-
-