Model Number PRD-SC30-MICRO
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 10/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Event or Problem Description
A stealth 360 peripheral orbital atherectomy device (oad) was used for a lower extremity peripheral endovascular intervention (pvi).A non-csi microcatheter was placed in the tibio-peroneal trunk and crossed a heavily calcified lesion with a viperwire advance guide wire.The physician unsuccessfully attempted to advance the microcatheter distally due to the lesion.However, the physician believed this was the vessel's true lumen and began treatment with the oad.During the first treatment on low speed, the oad suddenly stopped.The physician attempted to retract the crown with the crown advancer knob but was unsuccessful.Glideassist mode was activated multiple times in an attempt to free the crown, but each time the oad stopped.Imaging revealed dye staining.The physician stopped the intervention, as there was a large collateral vessel filling the tibio-peroneal trunk that had not been compromised.The physician manually pulled the oad out of the tibio-peroneal trunk and tissue wrapped around the oad crown was observed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTH 360 ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55125
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key15772283
Report Number3004742232-2022-00259
Device Sequence Number20116779
Product Code MCW
UDI-Device Identifier10850000491264
UDI-Public(01)10850000491264(17)240731(10)441544-1
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberPRD-SC30-MICRO
Device Catalogue Number7-10059-01
Device Lot Number441544-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Initial Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-