| Model Number |
INFUSOR, PRESSURE, 500 ML, NETTED, 12/CS |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to vyaire medical that the infusor, pressure, 500 ml, netted, 12/cs has an air leakage during patient's surgery.The customer confirmed no harm was done to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation:the suspect device lot number pass all the inspection requirement on function, visual, and packaging requirement.The complaint's air leakage was considered as supplier material issue of tpu film.Therefore, the defect reported by the customer was confirmed.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|