| Model Number |
26816 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the device was not reading correctly.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the device was not reading correctly.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacuring narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
|
| |
|