U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CAPIOX RX25 OXYGENATOR WEST; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CAPIOX RX25 OXYGENATOR WEST; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX25RW
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the manifold malfunctioned; the luer connection at the proximal end before the first stopcock came apart.No patient involvement.Product was changed out.Procedure completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 10, 2022.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h4 (device manufacture date) h6 (identification of evaluation codes 11, 3331, 4114, 3259, 4307) type of investigation #1: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #2: 3331 - analysis of production records type of investigation #3: 4114 - device not returned investigation findings: 3259 - improper physical structure investigation conclusions: 4307 - cause traced to component failure the affected sample was not returned for evaluation; however, a picture was provided.The picture confirmed that the lock adaptor was not attached to the manifold.However, the lock adaptor is a press fit and should be able to be pushed back onto the manifold.A representative retention sample was inspected for damage with no anomalies noted on the device specifically with the manifold lock adaptor.All manifold units are 100% visually inspected and leak tested during the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX25 OXYGENATOR WEST
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key15774669
Report Number1124841-2022-00171
Device Sequence Number12426440
Product Code DTZ
UDI-Device Identifier00699753450110
UDI-Public(01)00699753450110
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153213
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number3CX*RX25RW
Device Catalogue NumberN/A
Device Lot Number2D21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer01/09/2023
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-