| Model Number |
AX2100BIF |
| Medical Device Problem Codes |
Crack (1135); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The following information was reported via medsun uf/importer report# (b)(4): "head lamp used by surgeon was noted to have cracked insulation/wire causing heat to build up (fire hazard).Noticed during plugging/unplugging for xray intra-op.The unit was switched out with new one.Test performed (b)(6) 2022; results: headlamp sent to biomed department for evaluation and correction.Have 2 new headlamps on order." no patient/user injury, death or surgical delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspected ultralite pro headlight w/premium cable (ax2100bif) was not returned for evaluation after three good faith efforts (gfe).The lot number information has not been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.A definitive root cause could not be determined.The issue of cracked insulation may be the result of rough handling, environmental damage, or wear.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|