• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I FULL OSSEOTITE¿ TAPERED IMPLANT 3.25 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I FULL OSSEOTITE¿ TAPERED IMPLANT 3.25 X 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number FNT3211
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient identifier unknown / not provided.Age at time of the event unknown / not provided.Gender unknown / not provided.Patient weight unknown / not provided.Additional device information unknown / not provided.Device manufacturer date unknown / not provided.
 
Event or Problem Description
Doctor reported that implant fractured at the body of the implant and had to be removed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).One (1) full osseotite tapered implant 3.25 x 11.5mm (fnt3211) was returned for investigation.Visual evaluation of the as returned product identified signs of use.Fracture identified at threads below platform of implant.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.A pre-existing condition noted on the per was unknown bone density type.The reported device was located on an unknown tooth site (fdi) and was used for approximately 8 months, 6 days.Dhr review was completed for the subject lot number (2018112225).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance's, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2018112225) for similar events and no other complaint was identified.Review completed utilizing keywords: fracture implant.Therefore, based on the available information, device malfunction did occur, and the reported event was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL OSSEOTITE¿ TAPERED IMPLANT 3.25 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15776124
Report Number0001038806-2022-01744
Device Sequence Number17374455
Product Code DZE
UDI-Device Identifier00844868003281
UDI-Public(01)00844868003281(17)231213(10)2018112225
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K063286
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/13/2023
Device Model NumberFNT3211
Device Catalogue NumberFNT3211
Device Lot Number2018112225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer03/04/2023
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-