| Model Number |
IPN915321 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "(b)(6) hospital used an iabp catheter in the catheterization laboratory.After the catheter was insert into a blood vessel and the machine was turned on to counter pulsate, the doctor noticed an air leak in the machine and observed blood in the helium connection tube at the end of the catheter, the doctor judged that the balloon had ruptured and then removed the catheter.A second iab was inserted at the same insertion site".The patient status is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported that "shanghai first people's hospital songjiang branch used an iabp catheter in the catheterization laboratory.After the catheter was insert into a blood vessel and the machine was turned on to counterpulsate, the doctor noticed an air leak in the machine and observed blood in the helium connection tube at the end of the catheter, the doctor judged that the balloon had ruptured and then removed the catheter.A second iab was inserted at the same insertion site".The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned for investigation was a 30cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) with the original packaging box that matches the serial number on the returned sample.The sample was returned in a cardboard box and was loosely packed within the original packaging box.Returned with the sample and connected to the iabc short driveline tubing was the supplied 30cc inflation driveline tubing; dried blood was noted within the inflation driveline tubing.Upon return, a 0.025in guidewire was noted inserted within the iabc central lumen.The one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.Dried blood was noted on the exterior surfaces and within the helium pathway.The customer photos were reviewed.In the photos, blood was noted within the helium pathway of the iabc bladder and inflation driveline tubing, which is consistent with the reported complaint.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve and it held for at least 1 minute and then 30 seconds five separate times.The 0.025in guidewire returned within the iabc central lumen was removed without difficulty; some dried blood was noted on the guidewire.The catheter's central lumen was successfully aspirated and flushed using a 60cc lab-inventory syringe.Some blood was noted.The iabc was leak tested.A leak was immediately detected from the outer lumen.Under microscopic inspection, a leak site consistent with contact from a sharp object was noted on the outer lumen approximately 24.5cm from the iabc distal tip.No other leaks were detected.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.Resistance was noted at approximately 68.3cm from the iabc distal tip.The guidewire was able to advance through the central lumen.No blood or debris was noted.The guidewire was front loaded through the iabc luer.Resistance was noted at approximately 9.2cm from the iabc luer.The guidewire was able to advance through the central lumen.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of iab blood in helium pathway is confirmed.During the investigation, the iab was confirmed with a leak from the outer lumen near the bladder, which caused blood to enter the helium pathway.The leak site identified is consistent with contact from a sharp object.The returned iabc bladder membrane was fully intact, with no leaks noted.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the leak from the outer lumen.The root cause of the outer lumen leak is undetermined.The most probable potential cause of how the catheter came into contact with a sharp object is customer handling.No further action required at this time.This will be monitored for any developing trends.
|
| |
|