EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
|
Back to Search Results |
|
| Model Number |
LOGIC CR TIB INSERT SLOPE+, SZ 4, 9MM |
| Medical Device Problem Codes |
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Aseptic loosening (4883)
|
| Date of Event |
10/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported an 80 yo female patient, initial right knee implanted on (b)(6) 2019, underwent a revision on (b)(6) 2022, approximately 3 years 8 months post the initial procedure.The full knee was revised.Loosening of femur due to significant poly wear (patella and tibia) were indicated.The patient was last known to be in stable condition following the event.No devices returning as they were disposed of by the hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 2822789, 02-010-03-0335, logic cr femoral cem, right, sz 3.5, 323871, 203-96-40, (11-3661) stryker sys 6 90x25x1.27, 332364, 203-96-42, (11-3671) stryker sys 6 90x13/21x1.19, 5279765, 201-78-15, holding pin mini sharp point 4 pk, 5284137, 201-78-81, 3 trocar, mod.Hex 2pk, 5284161, 201-78-81, 3 trocar, mod.Hex 2pk, 5284175, 201-78-81, 3 trocar, mod.Hex 2pk, 5404146, 02-012-45-3535, lgc tibial fit tray cem sz 3.5f / 3.5t, 6038382, 200-02-41, three peg patella 41mm, s010942, 521-78-36, threaded pin size 5.1 collarless, s018337, 521-78-23 , threaded pin size 2.3 collared, s018357, 521-78-23, threaded pin size 2.3 collared.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitants: 02-010-04-0340 - logic cr femoral por, right, sz 4, (b)(6).02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t, (b)(6).200-02-41 - three peg patella 41mm (b)(6), b5, d4, d6a, d10, h4.
|
| |
|
Event or Problem Description
|
|
Initial right knee implanted on (b)(6), 2018, underwent a revision on (b)(6), 2022, approximately 4 years 8 months post the initial procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
After review of additional information received the following sections, have been updated accordingly: h6: the revision reported was likely the result of femoral loosening and prosthesis wear of the tibial insert and patella component.The wear may have been due to a combination of third body wear, positioning of the components, and/or patient related conditions.The reported loosening may have been due to an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the femoral component.However, this cannot be confirmed because the components were not returned for evaluation and radiographs were not provided.It is unclear if the reported loosening may have contributed to the wear or if the wear may have contributed to the reported loosening.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b4, d1, d4, h6 -health effect - clinical code, medical device problem code, component code, type of investigation, investigation conclusion and h11.The revision reported was likely the result of femoral loosening and prosthesis wear of the tibial insert and patella component.The wear may have been due to a combination of third body wear, positioning of the components, and/or patient related conditions.The reported loosening may have been due to an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the femoral component.However, this cannot be confirmed because the components were not returned for evaluation and radiographs were not provided.It is unclear if the reported loosening may have contributed to the wear or if the wear may have contributed to the reported loosening if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
This is report 1 of 3 for this event/patient.
|
| |
|
Search Alerts/Recalls
|
|
|