• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDVRA500Q
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2022
Event Type  malfunction  
Event Description
It was reported that the implantable cardioverter defibrillator delivered appropriate shock therapy.It was then noted that the device exhibited charge time out error.Capacitor maintenance was performed and charge times were then normal.The device converted the patient successfully.No further intervention was performed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15777447
MDR Text Key306265646
Report Number2017865-2022-45007
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberCDVRA500Q
Device Catalogue NumberCDVRA500Q
Device Lot NumberS000078125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/28/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-