• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERYAN MEDICAL BIOMIMICS 3D VASCULAR STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERYAN MEDICAL BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Model Number 142122-10
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Restenosis (4576)
Date of Event 09/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
This mdr report is related to mdr number 3011632150-2022-00096.The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2022.The subject was implanted with two biomimics 3d (bm3d) stents (one 6.0 x 60mm stent and one 6 x 100mm stent) to treat a denovo lesion of the sfa proximal third to middle third and sfa middle third to distal third artery in the right leg.A retrograde approach was used and pre- and post-dilatation was conducted with percutaneous transluminal angioplasty (pta) and the lesion was prepared using atherectomy.On the (b)(6) 2022, an event of restenosis of treated segment (target lesion) was reported from the site.It was reported as "possibly related" to the device and "not related" to the procedure.It was also reported as target lesion related.The patient attended for an office visit, complaining of recurrent claudication symptoms in her right leg.An ultrasound showed concern for recurrent stenosis within the distal right superficial femoral artery.On (b)(6) 2022, the patient had an angiogram which revealed two areas of critical stenosis at the proximal end of the proximal right sfa stent as well as at the distal end of the distal sfa stent and a critical stenosis in the mid popliteal artery.There was soft thrombus lining the previously placed stents as well as the mid popliteal artery.Laser atherectomy and standard balloon percutaneous transluminal angioplasty (pta) was done on all critical stenoses with good results.Completion imaging revealed wide patency of the sfa and popliteal arteries.The intervention was reported as a target lesion revascularisation.The devices remain implanted.The patient outcome is reported as "resolved/recovered".
 
Additional Manufacturer Narrative
This is related to mdr number 3011632150-2022-00096.There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of claudication/pain and restenosis leading to intervention are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
This is related to mdr number 3011632150-2022-00096.There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of claudication/pain and restenosis leading to intervention are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.Section b.5.Was updated, section d.10.Was updated to include aspirin, sections d.3.And g.1.Were updated with veryan's site address, sections g.6.And h.2.Were updated with the type of report (follow-up 01) and the reason and section h.11.Reflects the sections changed in this report.
 
Event or Problem Description
This mdr report is related to mdr number 3011632150-2022-00096.The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2022.The subject was implanted with two biomimics 3d (bm3d) stents one 6.0 x 60 mm stent and one 6 x 100 mm stent (the subject of this report) to treat denovo lesions of the sfa proximal third to middle third and sfa middle third to distal third arterial segments in the right leg.A retrograde approach was used and pre- and post-dilation was conducted with percutaneous transluminal angioplasty (pta) and the lesions were prepared using atherectomy.On (b)(6) 2022 a restenosis of treated segment (target lesion) was identified by the site.It was reported as "possibly related" to the device and "not related" to the procedure.It was also reported as target lesion related.The patient attended for an office visit, complaining of recurrent claudication symptoms in her right leg.An ultrasound showed concern for recurrent stenosis within the distal right sfa.On (b)(6) 2022, the patient had an angiogram which revealed two areas of critical stenosis at the proximal end of the proximal right sfa stent as well as at the distal end of the distal sfa stent and a critical stenosis in the mid popliteal artery.There was soft thrombus lining the previously placed stents as well as the mid popliteal artery.Laser atherectomy and standard balloon percutaneous transluminal angioplasty (pta) were done on all critical stenoses with good results.Completion imaging revealed wide patency of the sfa and popliteal arteries.The intervention was reported as a target lesion/vessel revascularisation (tlr/tvr).The devices remain implanted.The patient outcome was reported as "resolved/recovered".Additional information was provided by the site updating the concomitant medication to include aspirin which was started on (b)(6) 2022 and was ongoing at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL
block 5
parkmore east business park
galway, H91 V 0TX
EI  H91 V0TX
Manufacturer Contact
alan mcdonagh
block 5
parkmore east business park
galway, H91 V-0TX
EI   H91 V0TX
MDR Report Key15778233
Report Number3011632150-2022-00097
Device Sequence Number4873712
Product Code NIP
UDI-Device Identifier05391526850459
UDI-Public(01)05391526850459(17)230702(11)211201(10)0000123388
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/02/2023
Device Model Number142122-10
Device Catalogue Number142122-10
Device Lot Number0000123388
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer04/23/2024
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN; CLOPIDOGREL (PLAVIX); CLOPIDOGREL (PLAVIX); HEPARIN
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexFemale
Patient Weight106 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-