Model Number 02.115.350
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Voluntary report received (mw5112632): it was reported that "my orthopedic dr.Used a depuy synthes plate attach the right distal radius to the ulnar after correction was made.The plate used broke, while the arm was still in the cast.There was a lot of unnecessary pain caused by this break and had to go through two more surgery's.The plate was removed and i have kept the broken plate in my possession.I looked for any recalls and could not find one for this product.My name is (b)(6).Fda safety report id # (b)(4)." concomitant device reported: unknown screws: trauma (part# unknown; lot# unknown; quantity# unknown).This report is for one (1) 2.4 va dorsal dstl rad t-pl/3h hd/5h sft.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e3: reporter is a patient.H6 the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment ((b)(4) email message 03nov2022).Visual analysis of the photo revealed that the 2.4 va dorsal dstl rad t-pl/3h hd/5h sft had broken across the middle section.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the 2.4 va dorsal dstl rad t-pl/3h hd/5h sft.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.4 VA DORSAL DSTL RAD T-PL/3H HD/5H SFT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15778295
Report Number8030965-2022-09624
Device Sequence Number4897529
Product Code HRS
UDI-Device Identifier10886982035325
UDI-Public(01)10886982035325
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K102694
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.115.350
Device Catalogue Number02.115.350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: TRAUMA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
Patient Weight53 KG
-
-