Model Number APB-4-12-3D-SS
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the coil's push wire was broken while inserting the coil into the catheter.Continuous flush was administered during the procedure.The damage was on the middle segment of the pusher.Coil separation/break and premature detachment occurred.The coil was removed from the patient by pulling out the push wire.There was no surgical or medicinal intervention required.The push wire was bent/broken and was not intentional.There was no friction or difficulty during testing or delivery.The physician did not reposition the coil.The physician did not attempt to detach the coil.The physician did not rotate the delivery pusher during the procedure. the reported device and any accessory devices were prepared as indicated in the ifu.No patient symptoms or further complications were reported as a result of this event.The patient was undergoing surgery for treatment of a saccular, unruptured left paraclinoid aneurysm with a max diameter of 5 mm and a 2 mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was moderate.Ancillary devices include a sofia 5fr guide catheter and a sl-10 microcatheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15779236
Report Number9617601-2022-00151
Device Sequence Number4874246
Product Code KRD
UDI-Device Identifier00847536008120
UDI-Public00847536008120
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2024
Device Model NumberAPB-4-12-3D-SS
Device Catalogue NumberAPB-4-12-3D-SS
Device Lot Number223092552
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Initial Report FDA Received Date11/11/2022
Date Device Manufactured10/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-