| Model Number |
133S-MX-14-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: tear in the seam, causing the device to deflate.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported unknown side tear in the seam, causing the device to deflate.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed, opening device on the edge of the insert to shell bond side assessed as unidentified (tear) opening (shell thickness within specification).Additional observations: white particles in device inner surface are observed.Wear abrasion is observed.Additional, changed, and/or corrected data: d1, d4, d9, g4, h3, h4, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported unknown side tear in the seam, causing the device to deflate.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported unknown side tear in the seam, causing the device to deflate.Device was explanted.
|
| |
|