Catalog Number 04P75-32
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2022, abbott point of care (apoc) was contacted by a customer who reported that i-stat1 analyzer sn 324825 emitted smoke from battery compartment while attempting to run controls.The product was replaced at no charge.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that rechargeable batteries were being used at the time of the event.Therefore the analyzer is unlikely to become hot to touch, failing safe.The product was replaced and is being returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on (b)(6) 2023.The customer reported that smoke was emitted from the battery compartment of analyzer s/n (b)(6) while attempting to run controls.Failure analysis confirmed the complaint and determined the cause to be a burnt tantalum capacitor found at location c4 on the analyzer main board.A rocketware search spanning three months revealed no similar incidents and no evidence of a trend.Over the past year, the actual number of incidents caused by reliability-related failures of tantalum capacitors was (b)(4), which is less than the expected (b)(4) obtained by the reliability calculations.Therefore, no corrective or preventive action is required as the threshold has not been tripped and no product deficiency was found.Rather, this was a malfunction, which was attributed to the failure of the tantalum capacitor in the c4 location.The customer was using a battery carrier of the red (fused) configuration as a power source.The product failed-safe.There is no reason to believe that a capacitor failure while using a red carrier is likely to cause or contribute to death or serious injury.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Common Device Name
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
FLEXTRONICS MANUFACTURING (SINGAPORE)
pte ltd
1 kallang place
singapore 33921 1
SN   339211
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15779424
Report Number2245578-2022-00158
Device Sequence Number17792434
Product Code CGA
UDI-Device Identifier00054749003716
UDI-Public00054749003716
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K001387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04P75-32
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer01/19/2023
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-