• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884711006E
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported the stylet would not come out of the feeding tube and the tube had to be removed from the patient.The cap on the end broke before it was retracted.
 
Additional Manufacturer Narrative
The device history record (dhr) review could not be performed because the lot number was not available.One sample was received for evaluation.The sample was visually inspected, and the reported condition of hub disassembled from the wire was confirmed; however, hub detachment can be caused by not activating the hydromer with enough fluid.The hydromer was activated, and the stylet was able to be removed without problem.The investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.The instructions for use provide the proper procedure for insertion of the feeding tube and extraction of the stylet.No action plan is required at this point.The manufacturing site will continue to monitor customer complaints and feedback notifications for adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 10FR 43IN W STYLET EN
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfiled MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15779432
Report Number9612030-2022-03461
Device Sequence Number15103498
Product Code KNT
UDI-Device Identifier10884521582903
UDI-Public10884521582903
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711006E
Device Catalogue Number8884711006E
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-