(b)(4).Event date: only event year known: 2022.Batch # 330a51.Investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample determined that the tcr10 reload was received with the right gripping surface broken off making the reload nonfunctional.The reload was fully loaded with staples with the swing tab in the locked position.No functional testing was performed due to the condition of the reload.The event reported was confirmed and it is related to improper use of the reload.The condition on the returned reload is consistent with an improper loading technique.When loading the cartridge in the device, make sure the cartridge is inside the channel track and proper loading.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|