LDR MÉDICAL IMPLANT MOBIC M ST 15X17 H5 US; MOBI-C CERVICAL DISC PROSTHESIS Back to Search Results
Catalog Number MB3575
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that a mobi-c implant detached from the peek cartridge while the surgeon was rotating the implant intraoperatively.It was removed and another mobi-c implant was used to complete the case.There was no impact to the patient.
 
Event or Problem Description
It was reported that a mobi-c implant detached from the peek cartridge while the surgeon was rotating the implant intraoperatively.It was removed and another mobi-c implant was used to complete the case.There was no impact to the patient.
 
Additional Manufacturer Narrative
Inspection: the product was not returned and no photos were provided, so an evaluation is unable to be performed.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to inserting the device off-axis, adjusting during insertion, not turning the handle correctly during insertion or attempting to adjust the positioning while inserting the implant.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT MOBIC M ST 15X17 H5 US
Common Device Name
MOBI-C CERVICAL DISC PROSTHESIS
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15780530
Report Number3004788213-2022-00091
Device Sequence Number15117699
Product Code MJO
UDI-Device Identifier03662663000154
UDI-Public(01)03662663000154(17)251101(10)5374404
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P110009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2025
Device Catalogue NumberMB3575
Device Lot Number5374404
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer04/17/2023
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date04/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-