| Catalog Number |
MB3575 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c implant detached from the peek cartridge while the surgeon was rotating the implant intraoperatively.It was removed and another mobi-c implant was used to complete the case.There was no impact to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c implant detached from the peek cartridge while the surgeon was rotating the implant intraoperatively.It was removed and another mobi-c implant was used to complete the case.There was no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspection: the product was not returned and no photos were provided, so an evaluation is unable to be performed.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to inserting the device off-axis, adjusting during insertion, not turning the handle correctly during insertion or attempting to adjust the positioning while inserting the implant.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|