Model Number FV701T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: a deformation of the outer housing of the prosa valve was observed through the visual inspection.The measurement of the plane parallelism could confirm that with a value of -0,076 mm the housing membrane is clearly outside tolerance (0 ± 0.02 mm).Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the prosa valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of over/ under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The results show that the valve operates within the accepted tolerance in the vertical position.Internal inspection: after dismantling of the valve, no deposits were found in prosa.Results for the sake of thoroughness and transparency, we would like to point out that a permeability test was already carried out after the revision at the hospital.Based on our investigation results, we can determine a deformation of the outer housing.The deformation had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a prosa (part # fv701t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operated in overdrainage and non-adjustability.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 49 years.Height: 180 centimeters (cm).Weight: 80 kilograms (kg).Gender: female.Same event as #3004721439-2022-00338.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSA VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15781509
Report Number3004721439-2022-00337
Device Sequence Number12412535
Product Code JXG
UDI-Device Identifier04041906133023
UDI-Public4041906133023
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model NumberFV701T
Device Catalogue NumberFV701T
Device Lot Number20030149
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Initial Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
Patient Weight80 KG
-
-