ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 19 CM RETR; CATHETER, HEMODIALYSIS, IMPLA Back to Search Results
Model Number IPN001348
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
The complaint is reported as: with these catheters, we observed a rigidity more important than usual on the "tubing" of the catheter which leads to a kink linked to the clamps.When the clamp is open (in a dialysis session), it happens that the patient moves and that this kink obstructs the passage of blood and therefore stops the blood pump.Additional information received 28oct2022: the kinked is linked to the clamp when the catheter is not used.The tubing being slightly rigid, the kink persists.If the patient moves slightly during the session, the tube pinches at the kink.The clamps are moved by the teams at each session, but this is not sufficient to avoid the problem.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The complaint is reported as: with these catheters, we observed a rigidity more important than usual on the "tubing" of the catheter which leads to a kink linked to the clamps.When the clamp is open (in a dialysis session), it happens that the patient moves and that this kink obstructs the passage of blood and therefore stops the blood pump.Additional information received 28oct2022: the kinked is linked to the clamp when the catheter is not used.The tubing being slightly rigid, the kink persists.If the patient moves slightly during the session, the tube pinches at the kink.The clamps are moved by the teams at each session, but this is not sufficient to avoid the problem.
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-L 15 FR X 19 CM RETR
Common Device Name
CATHETER, HEMODIALYSIS, IMPLA
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15784846
Report Number9680794-2022-00759
Device Sequence Number14231443
Product Code MSD
UDI-Device Identifier10801902127171
UDI-Public10801902127171
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141051
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN001348
Device Catalogue NumberCS-15192-VFE
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2022
Supplement Date Received by Manufacturer12/07/2022
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEMODIALYSIS; HEMODIALYSIS
Patient SexUnknown
-
-