MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5530-G-611
Medical Device Problem Codes Degraded (1153); Unstable (1667)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 10/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Knee revision due to instability.
 
Event or Problem Description
Knee revision due to instability.
 
Additional Manufacturer Narrative
Reported event: an event regarding instability and wear involving a triathlon insert was reported.Instability was not confirmed, however, analysis of the returned device confirmed wear of the liner.Method & results: product evaluation and results: visual inspection: visual inspection of the returned device indicated that the insert is worn.The damage present on the posterior portion was consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.Dimensional inspection: dimensional inspection was not performed because the device was returned damaged.Functional inspection: functional inspection was not performed because the device was returned damaged.Material analysis: not performed as visual inspection provided no indication that the event was a result of a material integrity issue.If further information becomes available additional testing will be performed.Clinician review: a review of the provided medical records by a clinical consultant indicated: narrative: there was little medical information and some translational issues.The patient appears to have had a tka on the right for medial compartment disease, although the x-rays were not labeled.Over a period of 12 years the polyethylene wore and may have fractured.The patient was described as having no instability, so the indications for the surgery outside of the xray changes were not conveyed.A revision procedure was performed with poly exchange and patellar resurfacing.Intraoperatively a stable construct was accomplished.No follow-up data were provided.Conclusion/assessment: revision knee arthroplasty with poly exchange and patellar resurfacing for ¿patella problems¿ and poly wear/fracture.Event confirmation: the event of a revision can be confirmed.Root cause: the root cause of the revision was polyethylene wear and potential fracture and reported patella problems.The root cause of the poly wear/fracture cannot be ascertained.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: as per the clinician review; revision knee arthroplasty with poly exchange and patellar resurfacing for ¿patella problems¿ and poly wear/fracture.Event confirmation: the event of a revision can be confirmed.Root cause: the root cause of the revision was polyethylene wear and potential fracture and reported patella problems.The root cause of the poly wear/fracture cannot be ascertained.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, dated pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key15785312
Report Number0002249697-2022-01610
Device Sequence Number3038369
Product Code MBH
UDI-Device Identifier07613327050592
UDI-Public07613327050592
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2015
Device Model Number5530-G-611
Device Catalogue Number5530-G-611
Device Lot Number1JK073
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer02/08/2023
Initial Report FDA Received Date11/11/2022
Supplement Report FDA Received Date03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age82 YR
Patient SexMale
Patient Weight74 KG
-
-