MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10013
Device Problems Intermittent Continuity (1121); Connection Problem (2900); Noise, Audible (3273); Unintended Application Program Shut Down (4032)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a cori assisted tka procedure, upon starting to cut on distal femur, a ¿system time out error: the robotic drill has been deactivated¿ occurred.After clearing the error, the real intelligence robotic drill was shaking and making a high-pitched noise, even though the foot pedal was not being pressed.When backing out of the cut screen, or touching the refine button, the noise would stop.Then, it was attempted to swap out the attachment, however upon loading burr, a ¿robotic drill critical error¿ was received, case was quit out and ¿an internal error: application expectedly exited¿ was received.Surgery was resumed after a non significant delay, with a back up device.Patient was not harmed.In addition, both before and after the case, the drill passed the kpc test, however it was noticed that when manually twisting burr, it felt jagged.
 
Manufacturer Narrative
The real intelligence robotic drill, part # rob10013, serial # (b)(6), used for treatment, was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.No issues were identified during functional evaluation, device performed as expected.A log file review was performed.The reported problem was confirmed.Review of the logs from 2022-10-18 shows that warnings for "system time out" were linked to an exposure motor stall error.A possible cause may have been that debris buildup led to the bur getting stuck, this may have resulted in a drill motor stall which created the system time out errors.Additionally, it may have contributed to the jagged feeling experienced when manually twisting the bur.Another possible cause may have been excessive torque was applied to the drill when burring, this would have resulted in the drill motor stall.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of the reported complaint.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Manufacturer Narrative
The real intelligence robotic drill, part # rob10013, serial # (b)(6), used for treatment, was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.No issues were identified during functional evaluation, device performed as expected.A log file review was performed.The reported problem was confirmed.Review of the logs from 2022-10-18 shows that warnings for "system time out" were linked to an exposure motor stall error.The most likely cause of this event is associated with a failure of the drill exposure motor due to the thermo-mechanical stress induced within the motor at the encoder and electrical noise on the console error status inputs to the drill exposure motor encoder.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A historical review concluded that the lot, serial number or part number reported in this event is related to a corrective/preventive action already implemented.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Continuous improvements have been made to the cori robotic drill and manufacturing processes to reduce drill disconnection error messages.These improvements consisted of: 1.A hardware update to the cori console to reduce noise on the internal electronics.2.An update to the cori system¿s software and firmware to improve the user experience when error messages are displayed, and 3.A hardware update to the cori drill to reduce mechanical stress on drill exposure the motor.The first two improvements are fully deployed.The third improvement is being deployed for new orders and as drills are returned for routine servicing.Also, smith+nephew is voluntarily performing a recall/field notification for the cori real intelligence robotic drill.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15785842
MDR Text Key305532421
Report Number3010266064-2022-00654
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES# 93620
Patient Sequence Number1
-
-