| Model Number |
71631070 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an internal fixation surgery, when attempting to insert the trigen low profile screw 5.0mm x 30mm through the targeter, the screw could not be inserted.The screw and the long hexdriver were removed after confirmed using imaging that the tip of the long hexdriver's inner rod broke off and the piece was remained inside the screw head.Surgery was resumed, after a 5 minutes delay, with a back-up screw and driver.Patient was not injured as consequence of this problem.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms the inner shaft portion of the device threaded tip broke off.All pieces of the device were returned.The screw mention in the complaint description was returned as well and the visual does confirm the tip of the device stuck inside the screw.The device shows significant signs of wear and usage.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|