DEPUY MITEK LLC US MICRO QUICKANCHOR PLUS PRELOADED WITH 4/0 ETHIBOND SUTURE AND C-1 NEEDLES; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 212866
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
This is report 1 of 2 forpc-(b)(4).It was reported by the sales rep in canada that during an unknown surgery on (b)(6) 2022 , it was observed that the micro quickanchor plus preloaded with 4/0 ethibond suture and c-1 needles device broke off.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the customer provided a photo of the complaint device.Upon visual inspection of the photo, the two devices are shown inside a transparent plastic bag, however it is not possible to determine if the devices have an issue, the needles and the anchors are not visible in the photo.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, and considering that the device was not tested, this complaint cannot be confirmed.The device is required for testing, the photo provided does not contain enough evidence to determine why the customer experienced the failure, hands on analysis should provide the required evidence to provide a root cause.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection, it is observed that the anchor, needles, and suture were not received along with the device.No anomalies were found in the tip and the rest of the device.This complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device and no non-conformances were identified.At this moment, the root cause for the issue experienced by the customer cannot be determined.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO QUICKANCHOR PLUS PRELOADED WITH 4/0 ETHIBOND SUTURE AND C-1 NEEDLES
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15786173
Report Number1221934-2022-03511
Device Sequence Number11528641
Product Code HWC
UDI-Device Identifier10886705002535
UDI-Public10886705002535
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K024115
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model Number212866
Device Catalogue Number212866
Device Lot Number8L58023
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer11/18/2022
01/12/2023
01/24/2023
Initial Report FDA Received Date11/13/2022
Supplement Report FDA Received Date11/21/2022
01/13/2023
01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-