• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 190610
Medical Device Problem Codes Mechanical Problem (1384); Reflux within Device (1522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008k2 hemodialysis (hd) machine had a saline bag backfill during set-up for use.A patient was not connected to the machine at the time of the incident.Additional information was requested however a response was not received.No parts were returned to the manufacturer for a physical evaluation.
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008k2 hemodialysis (hd) machine had a saline bag backfill during set-up for use.A patient was not connected to the machine at the time of the incident.Additional information was requested however a response was not received.No parts were returned to the manufacturer for a physical evaluation.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation determined that there was no causal relationship between the objective evidence and the alleged event; therefore the alleged event is unconfirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key15788076
Report Number0002937457-2022-01954
Device Sequence Number5578684
Product Code KDI
UDI-Device Identifier00840861100859
UDI-Public00840861100859
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K994267
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number190610
Device Catalogue Number190610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/12/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-