ALCON RESEARCH, LLC - HUNTINGTON CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CCWTT3
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non healthcare professional reported that there was central divot in the middle of the lens.There was patient contact.Additional information has been received and lens was removed and replaced with no patient harm.
 
Additional Manufacturer Narrative
All product and batch history records are quality reviewed prior to product release.A qualified associated cartridge and viscoelastic were indicated.The file indicated the use of a non-qualified handpiece.The only qualified handpiece for the company lens is the company handpiece.The root cause for the reported complaint appears to be related to failure to follow the instruction for use (ifu).The instruction for use (ifu) instructs that a company qualified delivery system combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company recommends using the qualified company delivery system or any other company qualified combination.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15788362
Report Number1119421-2022-02410
Device Sequence Number12157291
Product Code MJP
UDI-Device Identifier00380652457453
UDI-Public00380652457453
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2025
Device Model NumberCCWTT3
Device Catalogue NumberCCWTT3.235
Device Lot Number15274126
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III CARTRIDGE D; MONARCH III INJECTOR; PROVISC OVD
Patient Age67 YR
Patient SexMale
-
-