| Model Number |
01.32.3648HCT |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
10/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 15 days after the primary surgery, the patient came in reporting pain due to a dislocation of the head from the liner and the cause is unknown.The surgeon revised the head, stem, and liner.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on (b)(6) 2022.Lot 2208324: 30 items manufactured and released on (b)(6) 2022.Expiration date: 2027-05-16.No anomalies found related to the problem.To date, 18 items of the same lot have been sold with no similar reported case during the period of review.Additional component involved: ball heads: mectacer 01.29.208 biolox delta ceramic ball head 12/14 ø 36 size s - 4 (k112115) lot 2119462: 50 items manufactured and released on (b)(6) 2022.Expiration date: 2027-03-13.No anomalies found related to the problem.To date, 47 items of the same lot have been sold with no similar reported case during the period of review.
|
| |
|