| Catalog Number |
190-000 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported 9 false positive results with id now covid-19 on 31oct2022.The patient was recommended to rest and take tylenol as needed.This mfr.Report addresses test one (1) of nine (9) tests.No additional patient information, including treatment and outcome, was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.A supplemental report will be provided after completion.H3 other text : device discarded, single use.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: e1 - country (united states).Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot m210013 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 190-000 / lot m210013, test base part number 190-430 / lot m210013.The lot met the required release specifications.A review of the complaints reported as false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot m210013 showed that the complaint rate is (b)(4).Based on the above summary, the investigation is deemed complete.The product will continue to be monitored and tracked.H3 other text : device discarded, single use.
|
| |
|
Event or Problem Description
|
|
The customer reported 9 false positive results with id now covid-19 on (b)(6)2022.The patient was recommended to rest and take tylenol as needed.This mfr.Report addresses test one (1) of nine (9) tests.No additional patient information, including treatment and outcome, was provided.
|
| |
|