• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP TRIAL INSERT NAV 22/55; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP TRIAL INSERT NAV 22/55; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 75000657
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
 
Event or Problem Description
It was reported that during thr, a part of the back lip of one (1) polarcup trial insert slotted 22/55 broke while placing the trials into the patient.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event or Problem Description
It was reported that during thr, a part of the back lip of one (1) polarcup trial insert nav 22/55 broke while placing the trials into the patient.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: it was reported that during total hip replacement, a part of the back lip of one (1) polarcup trial insert slotted 22/55 broke while placing the trials into the patient.The device intended for use in treatment was returned for investigation.The product was returned completely.Upon visual inspection, the reported failure mode could be confirmed, as 2 pieces at the distal tips were broken off.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The performed complaint history review for the complaint device revealed no previous complaints for the same batch, and a review for the part number over the past 12 months in scope with similar failure mode revealed 3 similar events.Previous investigations demonstrated that the performance of the device is within the risk levels that are anticipated in the risk management documentation.As part of the continuous improvement, smith + nephew is working on a design enhancement.No further escalation is required.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.Based on the performed investigations, the relationship between the reported event and the device was confirmed, and the root cause stays undetermined after investigation.Normal wear and tear are known to contribute to the reported event.Additionally, repeated steam sterilization processes could contribute to an embrittlement.To date, further actions are not deemed necessary.There is no objective indication that the reported devices failed to meet manufacturing specifications upon release for distribution.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.The need for corrective action is not indicated.Nevertheless, smith+nephew will continue to monitor this device for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP TRIAL INSERT NAV 22/55
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15788810
Report Number9613369-2022-00591
Device Sequence Number5298591
Product Code LXH
UDI-Device Identifier07611996112511
UDI-Public07611996112511
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/02/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75000657
Device Catalogue Number75023362
Device Lot NumberB69366
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer11/22/2022
12/30/2022
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date12/01/2022
01/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/16/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-