| Model Number |
HUM LNR IMPACT TIP 38MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, during impaction, the impaction tip broke into two (2) pieces when impacted.There was no surgical delay/prolongation.Sales rep was unable to obtain photos/x-rays.The device will be return.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) based on historical complaint investigations and the reported event, the fractured humeral liner impactor tip may have been the result of a stress riser introduced during impaction, which led to crack initiation, propagation, and ultimate fracture.Additionally, the device may have experienced material degradation if sterilized beyond the recommended parameters listed in the reprocessing instruction which could have led to the observed failure.The most probable root cause associated with the reported event of ¿broken / non-functional¿ is associated with a partial or full-thickness crack in the device materials.Section d9: device available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|