| Model Number |
D087031 |
| Medical Device Problem Codes |
Positioning Failure (1158); No Display/Image (1183); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Innovative health, llc became aware on 13-oct-2022 of a report from (b)(6) on a viewflex xtra ice device that would not steer and produced a completely black image during a procedure.Innovative health received the device for evaluation on 17-oct-2022.Upon investigation, the distal tip of the device was found to be fractured.The tip did not fall off completely and was attached to the shaft.Furthermore, the initial report of a steering failure and imaging failure were confirmed.No injury was reported.
|
| |
|
Event or Problem Description
|
|
This device was reported as not being able to steer, and the image went black during the procedure.Investigation of the returned device identified a fractured tip.No patient injuries were reported.
|
| |
|